5131140 – FLUKE 725 Multifunction Process Calibrator -FLUKE

SKU: 61224beff58f Category: Brand:

Description

Category: Process Calibrator
Product Name: FLUKE 725 Multifunction Process Calibrator
Brand: FLUKE
Packing Condition: Packed

Additional information

Weight1.625 kg
CATEGORY

RANGE

Multi

VOLTAGE

30 V

CURRENT

24 mA

FREQUENCY

1.00 to 10.00 kHz

RESISTANCE

3200 Ohm

TEMPERATURE

Multi

MODEL

FLUKE-725/APAC/EMEA

BRAND

Specification

Specifications: Fluke 725 Multifunction Process Calibrator

Measurement Accuracy
Voltage DC30.000 V0.02% + 2 counts (upper display)
30.000 V0.02% + 2 counts (lower display)
100.00 mV0.02% + 2 counts
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA0.02% + 2 counts
Resistance0.0 to 400.0 Ω0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency2.0 to 1000.0 CPM0.05% + 1 count
1.0 to 1100.0 Hz0.05% + 1 count
1.00 to 10.00 kHz0.05% + 1 count
Sensitivity1 V peak-to-peak-minimum
PressureAccuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC100.00 mV0.02% + 2 counts
10.000 V0.02% + 2 counts
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA (source)0.02% + 2 counts
24.000 mA (simulate)0.02% + 2 counts
Resistance15.0 to 400.0 Ω0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω1 Ω (excitation current 0.05 to 0.4 mA)
Frequency2.0 to 1000.0 CPM0.05%
1.0 to 1100.0 Hz0.05%
1.00 to 10.00 kHz0.25%
Waveform5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracyNI-1200.2°C
PT-100 (385)0.33°C
PT-100 (393)0.3°C
PT-100 (JIS)0.3°C
PT-200 (385)0.2°C
PT-500 (385)0.3°C
PT-1000 (385)0.2°C
Resolution0.1°C
J0.7°C
K0.8°C
T0.8°C
E0.7°C
R1.8°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
ResolutionJ, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK0.6°C
BP1.2°C
Source accuracyNI-1200.2°C
PT-100 (385)0.33°C
PT-100 (393)0.3°C
PT-100 (JIS)0.3°C
PT-200 (385)0.2°C
PT-500 (385)0.3°C
PT-1000 (385)0.2°C
Resolution0.1°C
Accuracy stated for 4-wire measurement.
J0.7°C
K0.8°C
T0.8°C
E0.7°C
R1.4°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
ResolutionJ, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK0.6°C
BP1.2°C
Specifications
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
RampsSlow ramp, Fast ramp, 25% step-ramp
Loop power functionVoltage24 V
Accuracy10%
Maximum current22 mA, short circuit protected
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Steps25% of range, 100% of range
Environmental Specifications
Operating temperature-10°C to +55°C
Storage temperature-20°C to 71°C
Operating altitude3000 m
Safety Specifications
Agency approvalsEN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size130 x 236 x 61 mm
Weight0.65 kg
Batteries4 AA alkaline batteries
Battery replacementSeparate battery compartment, accessable without breaking calibration seal
Side port connectionsPressure module connector, also used for remote real-time programming

Download

Reviews

There are no reviews yet.

Be the first to review “5131140 – FLUKE 725 Multifunction Process Calibrator -FLUKE”

Your email address will not be published. Required fields are marked *